High Resolution Dielectric / Impedance Analyzer ALPHA
|
|
Product
Description
The Alpha
dielectric / impedance analyzer combines a high class precision
broadband dielectric material characterization system with a
typical impedance analyzer for the low and middle impedance range.
This combination makes it an unique and easy to use instrument
with exceptional overall performance.
The Alpha analyzer covers 16 orders of magnitude impedance range,
12,5 orders of magnitude frequency range and 2m° basic phase
accuracy in a single instrument. This combination makes the it a
powerful tool which sets the pace for electric analysis
instrumentation. As a result, the Alpha can measure the
permittivity, conductivity and impedance of nearly all materials
or components broadband and over a wide temperature range without
changing sample geometry or sample cells.
|
 |
|
Design Features
Short specification
|
Frequency range (f):
|
3 µHz ... 10 MHz (12.5 decades)
|
|
Frequency accuracy:
|
10 ppm, 32 bit resolution
|
|
Impedance range (R, Z, X):
|
10 mOhm ... 100 TOhm (16
decades) |
|
Capacitance range (C):
|
1 fF ... 1 F (15 decades)
|
|
Phase angle accuracy (Ø):
|
0.002 °
|
|
Loss factor accuracy (D):
|
0.00003
|
|
AC signal level:
|
1 mV ... 3 Vrms (1 mV ... 106
Vrms with optional
voltage booster)
|
|
max. current: |
70 mA
|
|
DC bias signal level
|
-40 VDC ... +40 VDC
(-150 VDC ... +150 VDC with optional
voltage booster)
|
|
Measuring modes:
|
auto ranging,
auto reference, fixed reference |
|
Calibration modes:
|
load, short,
open, internal self calibration and diagnostics |
|
Input connections:
|
BNC, 3 terminals
|
|
Display: |
bright numerical display with
online status and value messages
|
|
Interface:
|
GPIB / IEEE488
|
|
Software:
|
WinDETA
data acquisition software with online graphics
|
|
|
|
|